Talboys Professional 5000I Incubating Orbital Shakers

from Troemner, LLC.

Talboys Professional 5000I Incubating Orbital Shakers from Troemner, LLC. 980472 980472-C 980473 980473-C

Patented Opti-Flow forced ventilation system.

  • Patented Accu-Drive shaking system.
  • Calibration mode for temperature and speed.

The Talboys Model 5000I Incubating Orbital Shakers are designed for a variety of shaking and incubating applications such as cell cultures.

Operating Features:

Accu-Drive Shaking System:
  • The exclusive patented Accu-Drive Shaking System delivers exceptional speed control, accuracy, and durability.
  • The system continuously monitors shaking speed and maintains set-point, even under changing loads.
  • When unit is set to a speed above 100rpm, the speed accuracy is ± 1% of set speed.
  • When set under 100rpm, speed accuracy is ± 1rpm.
Opti-Flow Forced Ventilation System:
  • The exclusive patented Opti-Flow Forced Ventilation System uses twin induction fans and air deflectors to deliver exceptional temperature uniformity and stability.
Microprocessor Control:
  • The variable speed microprocessor control provides consistent uniform shaking action.
  • Microprocessor will display last set-point and will restart if power is interrupted.
PID Temperature Controller:
  • Maintains precise temperature control from ambient +5°C to 65°C.
  • Easy-to-use controls allow user to adjust temperature in 1°C increments.
Triple Eccentric Drive:
  • Permanently lubricated ball bearings and maintenance free, brushless DC motor provide reliable service and continuous duty operation.
LED Display:
  • Touch pad controls with easy-to-read, independent LED displays for temperature, speed, and time allow operator to view all settings at once.
  • Provides repeatable and accurate results every time and is easily visible across lab benches.
  • The timer will display elapsed time or, when programmed to user defined limit, will shut off unit when time reaches zero.
  • Display will show last settings, even after power has been turned off.
Temperature Calibration Mode:
  • Allows user to calibrate unit to an external temperature device.
Speed Calibration Mode:
  • Allows user to automatically recalibrate speed display.
RS-232 Interface:
  • Provides two-way communication for data logging and unit control.

Safety Features:

Load Sensor:
  • A built-in load sensor detects unbalanced conditions and automatically reduces rpm to a safe speed, or stops unit, to protect samples.
Overload Protection:
  • Audible and visual signals will activate when system detects an obstruction or overload of the tray.
Overshoot Protection:
  • Unit will shut down and activate audible and visual alarms if temperature deviates ± 1°C from set-point.
Speed Ramping Feature:
  • Slowly increases speed to desired set-point to avoid splashing.
Audible Alarm:
  • In timed mode, alarm will sound when the time reaches zero.
  • Alarm has optional mute function which can be set from the touch pad controls.
Polycarbonate Lid:
  • Permits viewing of samples without disturbing internal temperature.
  • When lid is opened, automatic shut off of the circuit will stop the shaker tray to protect operator.
Spill Resistant Design:
  • Channels fluids away from internal components.
Caution Hot Indicator:
  • Symbol illuminates when the temperature of the air in the chamber reaches 40°C and remains lit until temperature cools down.

Operating Conditions:

  • Units can be run in conditions from 15 to 32°C (59 to 90°F), maximum 80% relative humidity, non-condensing.

Applications:

  • Cell cultures, solubility studies, bacterial suspensions, and extraction procedures.

Ordering Information:

  • Unit includes a 92" (234cm) detachable, 3-wire cord and plug (230 volt units are supplied with Euro type plug).
  • Unit is also supplied with a non-skid rubber mat that fits the 18 x 18" (45.7 x 45.7cm) tray.
  • 5 year limited warranty on parts and labor.

Optional NIST Traceable Calibration Certificate.

  • Provided by Troemner's ISO / IEC 17025 accredited laboratory.
  • Multiple data points within the temperature and speed ranges, and time function, are reported on the certificate with the associated uncertainties.

PRODUCT TECHNICAL DATA SHEET